MODEL AND SUFFIX CODES
Model AAI143 Analog Input Module (4 to 20 mA, 16-channel, Isolated)
Suffix
Codes
-S Standard type
-H With digital communication (HART protocol)
5 With no explosion protection
E With explosion protection
0 Basic type
3 With ISA Standard G3 option and temperature (-20 to 70 °C) option
Option
Codes
/K4A00 With KS Cable Interface Adapter (For connecting AEA4D Terminal Board) [Model: ATK4A-00]
/A4S00 With Pressure Clamp Terminal Block [Model: ATA4S-00]
/A4S10 With Pressure Clamp Terminal Block (surge absorber) [Model: ATA4S-10]
/A4D00 With Dual Pressure Clamp Terminal Block [Model: ATA4D-00]
/A4D10 With Dual Pressure Clamp Terminal Block (surge absorber) [Model: ATA4D-10]
/CCC01 With Connector Cover for MIL Cable [Model: ACCC01]
Current Input Modules (Isolated)
This module provides 16 inputs of 4 to 20 mA signal. It can be used in dual-redundant configuration.
Items Specifications
Model AAI143 (*1)
Number of input channels 16, isolated
Input signal 4 to 20 mA
Allowable input current 24 mA
Withstanding voltage Between input and system: 1500 V AC, for 1 minute (*4)
Input resistance Power ON 270 Ω (20 mA) to 350 Ω (4 mA) (*2)
Power OFF 500 kΩ or larger
Accuracy ±16 µA
Data update period 10 ms
Transmitter power supply 19.0 V or higher (at 20 mA) 25.5 V or less (at 0 mA) (output current limit: 25 mA) (*5)
Setting of 2-wire or 4-wire transmitter For each channel by setting pin
Drift due to ambient temperature change ±16 µA/10 ˚C
Maximum current consumption 230 mA (5 V DC), 540 mA (24 V DC)
Weight 0.3 kg
External communication Pressure clamp terminal, MIL connector cable, dedicated cable (KS1)
HART communication (*3) Available
We are specialized in intelligent instruments, including : Transmitters, Flowmeters, Analyzers,Communicators, Safety Barriers, Sensors,Gas Detectors, PLC(DCS), Multimeters, Calibrators, Communicators, Recorder, Inverters, Positioners, Valves, Soft Starters, Circuit Breakers, Contactors, Rail Switches, Power Supply, HMI, Terminal Block, Relay etc.