MODEL AND SUFFIX CODES
Description
Model SAI143 Analog Input Module (4 to 20 mA, 16 channels, module isolation)
Suffix
Codes
-S Standard type
-H With HART communication [Release: R1.02.00] (*1)
5 For pressure clamp terminal block or MIL cable with no explosion protection
6 With signal cable interface adapter and no explosion protection (*2)
E For pressure clamp terminal block or MIL cable with explosion protection
F With signal cable interface adapter and explosion protection (*2)
3 With ISA Standard G3 and temperature (-20 to 70 ºC)
C With ISA Standard G3 and high-density installation (-20 to 70 ºC) (*5)
Option
Codes
/A4S00 With pressure clamp terminal block for analog
[Model: STA4S-00]
/A4S10 With pressure clamp terminal block for analog (with surge absorber)
[Model: STA4S-10]
/A4D00 With dual-redundant pressure clamp terminal block for analog (*3)
[Model: STA4D-00]
/A4D10 With dual-redundant pressure clamp terminal block for analog (with surge absorber) (*3)
[Part name: STA4D-10]
/PRP With prevention pin of false insertion (*4)
/CCC01 With connector cover for MIL cable (for flat ribbon cable)
[Model: SCCC01]
Description
STANDARD SPECIFICATIONS
l Current Input Module
This module inputs 4 to 20 mA signals from the field.
The SAI143 can be made dual redundant.
Item Specifications
Model SAI143 (*1)
Number of input channels 16 channels, module isolation (*5)
Rated input range 4 to 20 mA
Permissible input range 0 to 25 mA
Input impedance during Power-ON 250 Ω+ Current value
Voltage drop in the input protection circuit inside the module (*6)
Input impedance during Power-OFF 500 kΩ minimum
Rated accuracy SAI143-S : ±16 µA (Input range 1 to 23 mA)
SAI143-H : ±16 µA (Input range 0.05 to 23 mA) (*3) (*4)
Data update frequency 40 ms
Temperature drift ±16 µA/10 °C
Transmitter power supply
16.15 V minimum (at 20 mA)
26.4 V maximum (at 0 mA)
(Output current limit: 25 mA)
2/4-wire setting Individual channel setting
Changed by setting pins
Withstand voltage 1.5 kV AC between input signal and system for 1 minute (*2)
Maximum current consumption 5 V system: 320 mA,
24 V system: 550 mA
Weight Approx. 0.34 kg (For pressure clamp terminal block or MIL cable)
Approx. 0.39 kg (With signal cable interface adapter)
External connection
Pressure clamp terminal
MIL cable
Dedicated signal cable
Support PRM Available (SAI143-H)
Transmitter power supply ON/OFF
function (*7) Available in 2-wire setting, for all 16 channels at the same time.
We are specialized in intelligent instruments, including : Transmitters, Flowmeters, Analyzers,Communicators, Safety Barriers, Sensors,Gas Detectors, PLC(DCS), Multimeters, Calibrators, Communicators, Recorder, Inverters, Positioners, Valves, Soft Starters, Circuit Breakers, Contactors, Rail Switches, Power Supply, HMI, Terminal Block, Relay etc.